| Model Number |
5440030 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from the healthcare professional (hcp) via the manufacturer representative regarding an event which occurred during a tlif (posterior lumbar surgery) in a patient diagnosed with lumbar 4/5 spinal stenosis.It was reported that the screw plugs were damaged.When screwing in the screw plug to press the rod, the thread was deformed.The date of implant and explant is (b)(6) 2021.Product will be returned, and it was replaced with a medtronic product.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: part # 5440030, lot # h5508607 visual and macroscopic inspection confirmed the threads of the set screw have been damaged.The thread crest and flank damage appears to have initiated at the start of the thread, and is consistent around the damaged portion of the thread.The break off portion is still attached and the female torx of the screw has not been damaged.This type of damage is consistent with misalignment of the set screw threads during construct assembly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|