THOMMEN MEDICAL AG CONTACT IMPL., PF 6.0, E 5.0, L 9.5; DENTAL IMPLANT Back to Search Results
Model Number 4.13.213
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 12/04/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Loss of osseointegration, implant achieved osseointegration, primary stability was achieved, implant was completely covered with bone, no thread tap used, peri-implantitis, mobility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTACT IMPL., PF 6.0, E 5.0, L 9.5
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH  2540
MDR Report Key11299675
Report Number3003184527-2021-00244
Device Sequence Number10959934
Product Code DZE
UDI-Device Identifier07640156471724
UDI-Public7640156471724
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K072649
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/04/2013
Device Model Number4.13.213
Device Lot Number5108
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2021
Initial Report FDA Received Date02/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-