• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB SERVO-I; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB SERVO-I; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number SERVO-I
Device Problems Leak/Splash (1354); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Date 01/21/2021
Event Type  malfunction  
Event Description
It was reported that there was a leakage noise from the gas module in the ventilator.There was no patient involvement.Manufacturer's ref #: (b)(4).
 
Manufacturer Narrative
The ventilator was investigated on site and the oxygen gas module was diagnosed to be the cause to the reported event.However the customer decided to repair the unit by them self/third part service.No log files have been received.If this fault happens during ventilation, the patient may receive an amount of oxygen in the gas mixture that deviate from the expected.Flow and pressure may also deviate from the expected.Alarms will be activated if the failure occurs during ventilation.As no goods or log files were returned for investigation, the cause of the reported failure could not be determined.
 
Event Description
Manufacturer's ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-I
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key11299899
MDR Text Key230991653
Report Number8010042-2021-00300
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K123149
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSERVO-I
Device Catalogue Number6487800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2021
Initial Date FDA Received02/09/2021
Supplement Dates Manufacturer Received01/21/2022
Supplement Dates FDA Received01/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-