| Model Number |
977D260 |
| Medical Device Problem Codes |
High impedance (1291); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported during trial procedure, that lead contacts 13, 14 <(>&<)> then 9 ( 5, 6 <(>&<)> 2) were broken when rep checked the lead connectivity during trial.They removed the leads from the wens and re-inserted it several times to make sure the contacts were properly seated.The leads were cleaned and reinstalled.They still showed up as do not use when the connectivity was checked. manufacturer representative (rep) had concerns it might be the wens, but they switched the leads inside the wens and the same contacts showed up is broken confirming that the issue was the lead and not the wens.Eventually a third contact also appeared to be broken.Rep used a different lead from trunk stock to replace the one in question.No patient symptoms were reported.Lead would be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.Analysis of the lead found no anomaly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|