• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7106-24
Device Problem Failure to Infuse (2340)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2020
Event Type  malfunction  
Event Description
Information was received indicating that a smiths medical pump was not infusing medical fluid.There was a suspected occlusion in the tubing.There was no patient injury and no reported adverse events.
 
Manufacturer Narrative
Other, other text: three pictures and one sample were returned for analysis.During testing of the physical sample, an occlusions was found between the filter and tube line.All mitigations on placed were verified and it was confirmed has been executing according, therefore no corrective actions could be implemented, it will be continue monitoring this failure condition in this product for threshold or escalation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
MDR Report Key11302008
MDR Text Key230982359
Report Number3012307300-2021-01055
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586023279
UDI-Public10610586023279
Combination Product (y/n)N
PMA/PMN Number
K974013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number21-7106-24
Device Lot Number4034003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2021
Was the Report Sent to FDA? No
Date Manufacturer Received04/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-