| Model Number |
551731 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
01/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.It should be noted that potential lot numbers were provided, as follows: lot 0061747995: production date: 09/21/2020; expiration date: 09/30/2021.Lot 0061732989: production date: 07/20/2020; expiration date: 06/01/2021.This event was reported to the fda under medwatch # 5098813.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: "upon removal of the perifix epidural catheter, a portion of the catheter tip was retained within the patient.The retained portion was left in place.".
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number 400501419.No sample was provided for evaluation.Further investigation of the complaint is not possible.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|