• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. IV ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. IV ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 354210
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Fall (1848)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as event 2 of b.Braun medical internal report number (b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
Event 2: as reported by the user facility: the device disconnected during use, causing leakage of blood and medication (normal saline, lactated ringers, and propofol).The device leaked between the primary line and extension set, and at the distal end of the extension set.Due to leakage onto the floor, a clinician slipped and fell.
 
Additional Manufacturer Narrative
Event 2: this report has been identified as b.Braun medical internal report number 400501290.A photograph was provided for further evaluation.The provided photograph is a diagram that highlights the affected area.Based on the data from the investigation we are unable to determine the root cause of the reported incident via the photo.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd
allentown 18109
MDR Report Key11302080
Report Number2523676-2021-00036
Device Sequence Number14660318
Product Code FPA
UDI-Device Identifier04046964182174
UDI-Public04046964182174
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K083723
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model Number354210
Device Catalogue Number354210
Device Lot Number00VL749523
Initial Date Received by Manufacturer 01/25/2021
Supplement Date Received by Manufacturer01/25/2021
Initial Report FDA Received Date02/09/2021
Supplement Report FDA Received Date03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-