• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAGER & MEISINGER GMBH INFINITY OCTAGON IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAGER & MEISINGER GMBH INFINITY OCTAGON IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number 504810BL
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This implant loss suggests that the source of the problem was likely to stem from no osseointegration and mobility.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, overall health or behavior.Proper intended use of the implant is described in its instructions for use.
 
Event or Problem Description
Asymptomatic, mobility and previous bone augmentation ((b)(6) 2020) were reported.Bone quality at the time of implant failure type iii.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability was achieved.Osseointegration was not achieved.Augmentation procedure (sinus lift) was performed at the time of surgery.Patient has high blood pressure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY OCTAGON IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
HAGER & MEISINGER GMBH
hansemannstrasse 10
neuss, germany
GM 
MDR Report Key11302558
Report Number1287163-2021-00798
Device Sequence Number741686
Product Code DZE
UDI-Device IdentifierE0HM504810BL1
UDI-PublicE0HM504810BL1
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K143539
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number504810BL
Device Catalogue Number504810BL
Device Lot NumberB75209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2021
Initial Report FDA Received Date02/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-