Model Number 2200-3010
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During inflation, the balloon ruptured below rbp.No patient injury reported, this mdr is being reported conservatively.Report source: foreign country: (b)(6).The angiosculpt device was discarded by the facility, thus no returned product investigation was performed.
 
Event or Problem Description
The angiosculpt device was used to treat a severely calcified and moderately tortuous mid lad.During inflation, the balloon ruptured at 14 atm.The procedure was completed with a different device.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT PTCA SCORING BALLOON CATHETER (RX)
Common Device Name
CATHETER, PTCA, CUTTING/SCORING
Manufacturer (Section D)
SPECTRANETICS
5055 brandin court
fremont CA 94538
MDR Report Key11303304
Report Number3005462046-2021-00009
Device Sequence Number3478814
Product Code NWX
UDI-Device Identifier00813132021207
UDI-Public00813132021207
Combination Product (Y/N)Y
Initial Reporter CountryTW
PMA/510(K) Number
P050018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2023
Device Model Number2200-3010
Device Catalogue Number2200-3010
Device Lot NumberG20010019
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient Weight73
-
-