| Model Number |
9438-06 |
| Medical Device Problem Code |
Premature End-of-Life Indicator (1480)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a low transmitter battery alert occurred.The product was evaluated.An external visual inspection was performed and passed.Voltage test was performed and failed due to 0 vdc.A review of the share logs was performed and a low transmitter battery alert was found within the investigation window.The allegation was confirmed.The probable cause was determined to be low transmitter battery voltage.In addition, a failed transmitter error was found in connection to the reported allegation.The reported issue of a low transmitter battery alert is reportable based on the finding of a failed transmitter error.No injury or medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|