• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE SYRINGE 0.5ML 31GA 8MM 10BAG 500 PL/WG; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE SYRINGE 0.5ML 31GA 8MM 10BAG 500 PL/WG; PISTON SYRINGE Back to Search Results
Model Number 928857
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
"multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 9231296.Medical device expiration date: 2024-09-30.Device manufacture date: 2019-08-19.Medical device lot #: unknown.Medical device expiration date: unknown.Device manufacture date: unknown.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.H.6.Fda device code: 1354, 1069.Fda patient code: 2687".
 
Event or Problem Description
It was reported that a syringe 0.5ml 31ga 8mm 10bag 500 pl/wg separated from the hub during use, and a syringe 0.5ml 31ga 8mm 10bag 500 pl/wg broke at the cannula during use.The following was reported by the initial reporter: "it was reported that the needle shield was hard to remove and one syringe that the needle hub separated when the shield was removed.Also, one syringe where the needle broke off into the consumer's stomach, but was able to remove the needle with tweezers.Verbatim: consumer reported found 5-6 syringes needle shield hard to remove consumer reported found 1 syringe shield hard to remove when removed the needle hub went with it.Lot #: 9231296 catalog#: 928857 date of event: (b)(6) 2021 needle hub removed with shield removal date of event unknown for the 5-6 syringes shields hard to remove samples status awaiting sample.Consumer reported 1 syringe when injected into stomach the needle broke off into stomach/site.She was able to remove with tweezers on her own.Denied reuse of needles.".
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2021-02-09.H6: investigation summary.Customer returned (1) 3/10cc syringe.Customer states that the needle broke off into the consumer's stomach.The returned syringe was examined and no broken cannula was observed.Customer returned (1) 1/2cc, 8mm, 31g walgreens syringe in an open poly bag from lot # 9231296.Customer states that the needle hub separated when the shield was removed.The returned syringe was examined and exhibited the hub-needle/shield assembly separated from the barrel.A review of the device history record was completed for batch# 9231296.All inspections and challenges were performed per the applicable operations qc specifications.There was one (1) notification noted that did not pertain to the complaint.There was one (1) notification noted for raised hubs.Root cause was unknown.Affected product was stated, contained, and destroyed.Capa1630423 has been opened to address this issue.
 
Event or Problem Description
It was reported that a syringe 0.5ml 31ga 8mm 10bag 500 pl/wg separated from the hub during use, and a syringe 0.5ml 31ga 8mm 10bag 500 pl/wg broke at the cannula during use.The following was reported by the initial reporter: "it was reported that the needle shield was hard to remove and one syringe that the needle hub separated when the shield was removed.Also, one syringe where the needle broke off into the consumer's stomach, but was able to remove the needle with tweezers.Verbatim: consumer reported found 5-6 syringes needle shield hard to remove consumer reported found 1 syringe shield hard to remove when removed the needle hub went with it.Lot #: 9231296.Catalog#: 928857.Date of event: (b)(6) 2021 needle hub removed with shield removal.Date of event unknown for the 5-6 syringes shields hard to remove.Samples status awaiting sample.Consumer reported 1 syringe when injected into stomach the needle broke off into stomach/site.She was able to remove with tweezers on her own.Denied reuse of needles.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 0.5ML 31GA 8MM 10BAG 500 PL/WG
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key11305412
Report Number1920898-2021-00178
Device Sequence Number14236024
Product Code FMF
UDI-Device Identifier00311917025360
UDI-Public00311917025360
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K024112
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number928857
Device Catalogue Number928857
Device Lot NumberSEE H.10.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2021
Initial Date Received by Manufacturer 01/20/2021
Supplement Date Received by Manufacturer03/01/2021
Initial Report FDA Received Date02/09/2021
Supplement Report FDA Received Date03/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-