OLYMPUS MEDICAL SYSTEMS CORP. VISERA PRO XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S40PRO
Device Problems No Display/Image (1183); Failure to Power Up (1476)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
A customer reported that the device couldn't turn on during a procedure.Reportedly, when the customer turned off the device and tried to turn it back on, the device did not turn on.Then, olympus inspected the device at the service department of olympus (b)(4) and found power supply and aperture failure.No endoscopic image was displayed on the monitor.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Other than the events reported in b5 of the initial mdr, there were no defects confirmed by device inspection by olympus korea co., ltd.Device history record review indicates that the device was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of the reported event could not be conclusively determined.However, olympus medical systems corp.(omsc) assumed that the reported event was caused by the defect of the power supply unit.This defect may have been caused by aged deterioration due to long-term use.If significant additional information is received, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA PRO XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11306136
MDR Text Key233829694
Report Number8010047-2021-02467
Device Sequence Number1
Product Code GCT
Combination Product (y/n)N
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 04/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLV-S40PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/18/2021
Initial Date FDA Received02/10/2021
Supplement Dates Manufacturer Received03/08/2021
Supplement Dates FDA Received03/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-