MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 50D; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 50D; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Material Deformation (2976); Difficult or Delayed Separation (4044)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the threads stripped on the g7 pps ltd acet shell 50d during the operation.After the surgeon had inserted the shell into acetabulum, he tried to remove the inserter.But couldn't separate the inserter from the shell, so he removed the shell and inserter together with other instruments.Then he separated the inserter from shell successfully in vitro and found the thread stripped and the metal chips had fallen off, doctor use 54 d shell complete the operation.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d4; d9; g3; h2; h3; h4; h6.  one g7 pps ltd acet shell 50d was returned and evaluated.Upon visual inspection the hex plugs have some scuffing visible but were able to be removed with the use of an allen wrench.When the porous coating is brushed some specks of the porous falls off no photos were taken of the specks as they were difficult to see.The threads of the apical hole are chipped.Dark debris is present within the apical plug.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.  multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2021 - 02305.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACET SHELL 50D
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11307591
Report Number0001825034-2021-00433
Device Sequence Number3256337
Product Code PBI
UDI-Device Identifier00880304524217
UDI-Public(01)00880304524217(17)281012(10)6395500
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number010000662
Device Lot Number6395500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/26/2021
Supplement Date Received by Manufacturer08/04/2021
Initial Report FDA Received Date02/10/2021
Supplement Report FDA Received Date08/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age47 YR
-
-