MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number SONATA
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user's hearing performance with the device was affected.In situ measurements showed affected channels.The user was re-implanted on (b)(6) 2021.
 
Additional Manufacturer Narrative
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Event or Problem Description
The user's hearing performance with the device was affected.In situ measurements showed affected channels.The user was re-implanted on (b)(6) 2021.
 
Event or Problem Description
The user's hearing performance with the device was affected.In situ measurements showed affected channels.The user was re-implanted on (b)(4) 2021.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However to determine an exact root cause a device investigation of the explanted device is necessary.The concerned device was explanted but has not been received for investigation yet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key11307670
Report Number9710014-2021-00095
Device Sequence Number3374466
Product Code MCM
UDI-Device Identifier09008737049188
UDI-Public(01)09008737049188
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberSONATA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/09/2021
Supplement Date Received by Manufacturer02/09/2021
02/09/2021
Initial Report FDA Received Date02/10/2021
Supplement Report FDA Received Date04/12/2021
06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age13 MO
-
-