ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE CD 13 MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported from the wir form that a persona knee instrument was returned and reported missing bearing.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Performed a visual inspection of the returned item and found it to exhibit signs of repeated use and has components disassembled /missing the components were not returned.The device history records & receiving inspection report were reviewed for deviations and/ or anomalies with no deviations/anomalies identified.Medical records were not provided.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE CD 13 MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11308002
Report Number0001822565-2021-00352
Device Sequence Number3256347
Product Code JWH
UDI-Device Identifier00889024245778
UDI-Public(01)00889024245778(10)63618903
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42527900303
Device Lot Number63618903
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/21/2021
Supplement Date Received by Manufacturer02/23/2021
Initial Report FDA Received Date02/10/2021
Supplement Report FDA Received Date02/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-