Model Number 1218-87-456
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Pain (1994); Physical Asymmetry (4573)
Event Date 06/04/2020
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Ad 15 jan 2021: unfiled claim and medical record received.After review of the medical records, the patient was revised to address painful, symptomatic modular metal on metal posterior total hip on the left side.Operative note reported of metallosis and osteolytic lesion behind the socket.The patient complained that the leg had been a little bit short.Doi: (b)(6) 2010, dor: (b)(6) 2020, left hip.
 
Manufacturer Narrative
Product complaint (b)(4) investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE MTL INS NEUT40IDX56OD
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11310955
MDR Text Key231326416
Report Number1818910-2021-02774
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2015
Device Model Number1218-87-456
Device Catalogue Number121887456
Device Lot Number3187008
Was Device Available for Evaluation? No
Date Manufacturer Received03/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
ASPHERE M SPEC 12/14 40 +1.5
Patient Outcome(s) Required Intervention;
Patient Age58 YR
-
-