• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL LIVEWIRE STEERABLE ELECTROPHYSIOLOGY CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL LIVEWIRE STEERABLE ELECTROPHYSIOLOGY CATHETER; CATHETER, STEERABLE Back to Search Results
Model Number 402032
Medical Device Problem Code Signal Artifact/Noise (1036)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One 6f, duo-decapolar, large curl, livewire ep catheter was received for evaluation.Electrodes 1-2, 9-10, and 13-14 displayed acceptable resistance values; however, a short circuit was detected between electrode 1 and conductor wire 13.Dissection revealed that the flat wire insulation was torn and the insulation for conductor wire 13 was abraded in the same location, consistent with the short circuit detected and the reported noise.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the short circuit is consistent with damage during use.
 
Event or Problem Description
This report is to advise of an event observed during analysis confirming a short circuit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIVEWIRE STEERABLE ELECTROPHYSIOLOGY CATHETER
Common Device Name
CATHETER, STEERABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
MDR Report Key11311250
Report Number2182269-2021-00010
Device Sequence Number3302589
Product Code DRA
UDI-Device Identifier05414734202817
UDI-Public05414734202817
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K022380
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number402032
Device Catalogue Number402032
Device Lot Number7442125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2021
Initial Report FDA Received Date02/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-