• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. HONEYCOMB; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. HONEYCOMB; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 71674075
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the honeycomb has cracked.The event was noticed by the sterile staff, after surgery, no surgical delay or injury to the patient was reported.
 
Additional Manufacturer Narrative
Results of investigation: the device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions will be taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HONEYCOMB
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11312296
Report Number1020279-2021-01121
Device Sequence Number14391176
Product Code LXH
UDI-Device Identifier03596010560711
UDI-Public03596010560711
Combination Product (Y/N)N
Initial Reporter CountrySW
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71674075
Device Catalogue Number71674075
Device Lot Number19FSP0005
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/18/2021
Supplement Date Received by Manufacturer03/02/2021
Initial Report FDA Received Date02/11/2021
Supplement Report FDA Received Date03/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-