• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT Back to Search Results
Model Number VENUL16120
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Expiry date: 10/2021.Device not returned.
 
Event or Problem Description
It was reported that during stent placement procedure, the stent allegedly failed to expand.There was no reported patient injury.
 
Event or Problem Description
It was reported, that during stent placement procedure in the left common iliac vein.The stent allegedly, failed to expand.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the lot history record of this lot was reviewed, with special attention to the manufacturing and inspection of this product.And the product was found to have met the specification prior to shipment.Investigation summary: the physical sample was not available, and images demonstrating stent or delivery system were not provided.The alleged issue could not be re produced, which led to an inconclusive evaluation result.Based on the information available, the investigation is closed with inconclusive result.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 10/2021).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT
Common Device Name
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
MDR Report Key11312932
Report Number9681442-2021-00030
Device Sequence Number2555628
Product Code QAN
UDI-Device Identifier00801741103667
UDI-Public(01)00801741103667
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P180037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVENUL16120
Device Catalogue NumberVENUL16120
Device Lot NumberANDX1445
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/18/2021
Supplement Date Received by Manufacturer04/13/2021
Initial Report FDA Received Date02/11/2021
Supplement Report FDA Received Date05/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-