| Model Number |
2760 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
01/21/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was reported that during interferential treatment, the customer allegedly received an overcharge and reported a second degree burn.A djo service and repair technician went onsite to perform an evaluation on the device.After performing functional and electrical testing, the technician was unable to duplicate the reported complaint.The product met manufacturer specifications.Possible root causes include moisture or misapplication of the electrodes.
|
| |
|
Event or Problem Description
|
|
It was reported that during interferential treatment, the customer allegedly received an overcharge and reported a second degree burn.
|
| |
|