• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB FLOW-I; GAS-MACHINE, ANESTHESIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB FLOW-I; GAS-MACHINE, ANESTHESIA Back to Search Results
Model Number FLOW-I C20
Device Problem Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/15/2021
Event Type  malfunction  
Event Description
It was reported that the anesthesia workstation had a constant leakage.There was no patient harm.Manufacturer´s ref #: (b)(4).
 
Manufacturer Narrative
The reported anesthesia system was investigated at the hospital.Received information states that the source of the leakages could be found in the manual ventilation membrane and in the volume reflector.The affected parts were replaced and after successful testing, the anesthesia system was returned for clinical use.The device logs, the replaced parts or additional information have not been available.We have not been able to determine the true cause of the reported issue.
 
Event Description
Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I
Type of Device
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key11314325
MDR Text Key231388427
Report Number8010042-2021-00352
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K191027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFLOW-I C20
Device Catalogue Number6677200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-