| Model Number |
VCP460H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot number qhmctj /vcp460w, and no non-conformances related to the reported complaint condition were identified.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: procedure name: an arrhythmia surgery.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ 275 g/m.Note: events reported via mw# 2210968-2021-01321.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an arrhythmia surgery on (b)(6) 2021 and suture was used.When tying the suture during continuous suturing, the weaving of the suture was torn except for center suture.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent to the fda: 3/23/2021.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional h3 investigation summary: the product was returned for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that one suture piece of product code vcp460h was received for evaluation and fraying, damaged, body fluids and marks on the surface that appears to be by use of the surgical instrument were noted.The product code vcp460 contains absorbable suture and as the sample was received open, the time of exposure of suture to the environment could not be determined and no functional test can be performed due to sample condition.The condition of the sample received indicates improper handling of the device.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, the device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent to the fda: 2/11/2021.Additional: g6.All manufacturer type of report: initial, 30-day the following information was reported in 2210968-2021-01319.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: please clarify if additional organ/tissue was dissected during needle removal ? no.Was the broken off needle fragment was retrieved? yes.Was broken needle part recovered during initial procedure or the patient had to return for re-operation back to or for removal of the needle? the broken needle part was recovered during initial procedure.Was there any patient consequences? please specify.No.Trade name - irgacare®, active ingredient(s) ¿ triclosan, dosage form ¿ suture/solid/parenteral, strength ¿ = 275 g/m.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent planned caesarean section on (b)(6) 2021 and suture was used.While the surgeon was suturing the uterine breach, the needle broke.To recover the missing part of the needle, the uterine breach had to be reopened and resutured with a new thread and needle.The needle was retrieved during the procedure and the procedure was completed successfully.There were no adverse patient consequences reported.
|
| |
|
Search Alerts/Recalls
|
|
|