BAXTER HEALTHCARE CORPORATION PRISMAFLEX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115269
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that during treatment with a prismaflex machine, the device had a missing blood pump crank.It was reported that blood was not returned to the patient a total of three times.The amount of blood loss was not reported.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Additional information: h3, h6 and h10.The monitor was evaluated on site by a qualified technician.Visual inspection found a missing crank on the device.The reported condition of missing crank was verified.To resolve the issue, the qualified technician replaced the missing blood pump crank.The cause of the condition could not be determined.The qualified technician performed prime, priming test and a 30 minute simulated treatment with no issues noted.The reported condition of blood loss was not verified.The cause of the condition could not be determined.A device history review revealed no issues that could have caused or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11315064
Report Number9616026-2021-00007
Device Sequence Number823723
Product Code KDI
UDI-Device Identifier07332414117702
UDI-Public(01)07332414117702(21)
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K131516
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number115269
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/18/2021
Supplement Date Received by Manufacturer03/17/2021
Initial Report FDA Received Date02/11/2021
Supplement Report FDA Received Date03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-