| Model Number |
Z9002 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
10/12/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age or date of birth, weight, ethnicity: unknown/not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis intraocular lens (iol) was explanted from the patient's operative eye due to unexpected post-op refraction.There was no additional surgical intervention required.Reportedly, the patient has recovered and no further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation: yes section d9: return to manufacturer: 2/25/2021 section h3: device evaluated by manufacturer: yes device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Haptic damage (bent) was observed, which can be attributed to handling and cannot be confirmed to be related to manufacturing.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|