| Model Number |
MX-410370 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: exchange from textured to smooth breast implants due to the patient¿s concern with the product and capsular contracture, baker grade iii/iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "exchange from textured to smooth breast implants due to the patient¿s concern with the product and capsular contracture, baker grade iii/iv".Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d.9, h.3, h.6.Device evaluation: deformation.A weight test of the device was verified and the device was within specification.Cloudy after autoclave disinfection process in the gel.Based on the device analysis the final assessment is: -no issues found related with the manufacturing.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "exchange from textured to smooth breast implants due to the patient¿s concern with the product and capsular contracture baker grade iii/iv".Device has been explanted.
|
| |
|