Concomitant medical products: product id: 8780, serial#: (b)(4), implanted: (b)(6) 2020, explanted: (b)(6) 2021, product type: catheter.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 22-nov-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
Information was received from a healthcare provider (hcp) via a company representative regarding a patient receiving baclofen (500 mcg/ml at minimum rate) via an implanted pump.It was reported that the patient reported to her managing doctor that she was not getting a therapeutic effect from the pump medication.The patient denied any trauma or falls.The managing doctor ordered a ct scan.The results of the scan showed that the pump had flipped, and it appeared that the catheter had disconnected from the pump.The patient was scheduled for a catheter revision on (b)(6) 2021.During the procedure, the catheter was found to be tangled and scarred into the pocket.The catheter was twisted all the way to the anchor in the spine.The catheter had broken off from the pump.The old twisted catheter was removed.A new catheter was implanted and reattached to the existing pump.The issue was noted to be resolved, and it was indicated that the hcp had no further information to provide regarding the event.
|