| Model Number |
TRANSDUCER X8-2T |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported an x8-2t model transducer had an articulation issue during use.There was no injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the transducer could not confirm the articulation issue as described by the customer.However, investigation of the device revealed the connector had corrosion and was chipped, a dented rack actuator, a worn i-tube, and deep bite marks on the window and tip shell.The transducer articulation mechanism was fully functional and met all articulation angle specifications.The physical damage to the transducer inhibited the overall performance of the device and is indicative of improper maintenance.
|
| |
|