• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: X7-2T; TRANSDUCER, ULTRASONIC, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

X7-2T; TRANSDUCER, ULTRASONIC, DIAGNOSTIC Back to Search Results
Model Number TRANSDUCER X7-2
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event or Problem Description
A customer reported an x7-2t model transducer had an articulation issue during use.There was no injury associated with this event.
 
Additional Manufacturer Narrative
Evaluation of the transducer confirmed the articulation issue as described by the customer.Functional testing of the device noted failure of the image and articulation tests.Visual inspection noted corrosion in the connector, a loose distal nut, and a scratched window.Destructive testing identified the steering cable pulled from the solder joint which resulted in the loss of articulation.The failed transducer was replaced by the philips service engineer at the customer site and no similar issues have been reported post repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X7-2T
Common Device Name
TRANSDUCER, ULTRASONIC, DIAGNOSTIC
MDR Report Key11322810
Report Number3019216-2021-10007
Device Sequence Number4124834
Product Code ITX
UDI-Device Identifier00884838061668
UDI-Public00884838061668
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRANSDUCER X7-2
Device Catalogue Number989605414122
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date02/02/2021
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/12/2021
Supplement Report FDA Received Date04/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
-
-