| Model Number |
10607 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.A 3.00 x 38 synergy drug-eluting stent was selected for use.However, as the protective sleeve was being removed, the shaft came apart around the rapid exchange point.The procedure was completed with another of same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy ii us mr 3.00 x 38mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of distal tip damage.A visual and tactile examination of the hypotube shaft found a shaft break 119.5 cm distal to the strain relief.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.A 3.00 x 38 synergy drug-eluting stent was selected for use.However, as the protective sleeve was being removed, the shaft came apart around the rapid exchange point.The procedure was completed with another of same device.There were no patient complications reported.
|
| |
|
Search Alerts/Recalls
|