| Model Number |
4003 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported that the sensors provided inaccurate pr values and intermittent spo2 readings.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed., other, other text: initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: bt12 6ba initial reporter phone # is entered as (b)(6) to meet the minimum allowable characters.Initial reporter phone # is: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported that the sensors provided inaccurate pr values and intermittent spo2 readings.No consequences or impact to patient were reported.
|
| |
|
Search Alerts/Recalls
|