SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 5 8MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
00584102508 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Our reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during tka surgery the zuk sz.5 8mm zuk poly trial was placed in the patient for sizing and broke in half inside.The procedure was completed without delay using a smith and nephew back-up device.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, used in treatment, was returned for evaluation.A visual inspection confirms the trial has broken into two pieces.Both pieces were returned.The device shows significant signs of wear/ usage.This instrument was manufactured in 2016.A medical investigation conducted and confirms this case reports the trial insert broke in half while being placed for sizing.A photo of the broken device confirms both pieces were retrieved.Per complaint details, there was no patient injury or delay reported and the procedure was completed using a backup device.Since no patient harm is alleged, no further clinical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|
|
|