Model Number 00584102508
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Our reference number: (b)(4).
 
Event or Problem Description
It was reported that, during tka surgery the zuk sz.5 8mm zuk poly trial was placed in the patient for sizing and broke in half inside.The procedure was completed without delay using a smith and nephew back-up device.No other complications were reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirms the trial has broken into two pieces.Both pieces were returned.The device shows significant signs of wear/ usage.This instrument was manufactured in 2016.A medical investigation conducted and confirms this case reports the trial insert broke in half while being placed for sizing.A photo of the broken device confirms both pieces were retrieved.Per complaint details, there was no patient injury or delay reported and the procedure was completed using a backup device.Since no patient harm is alleged, no further clinical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE PROVISIONAL SZ 5 8MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11325744
Report Number1020279-2021-01185
Device Sequence Number14484568
Product Code HSX
UDI-Device Identifier00885556619568
UDI-Public00885556619568
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00584102508
Device Catalogue Number00584102508
Device Lot Number63313529
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2021
Initial Date Received by Manufacturer 01/19/2021
Supplement Date Received by Manufacturer03/10/2021
Initial Report FDA Received Date02/12/2021
Supplement Report FDA Received Date03/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-