| Model Number |
5MAXACE068KIT |
| Medical Device Problem Codes |
Stretched (1601); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned ace68 confirmed that the device was stretched and ovalized on its distal shaft.This damage is typically a result of retracting the device against resistance.The root cause of the resistance could not be determined.Further evaluation of the device revealed a distal ovalization.This damage was likely incidental to the complaint.During functional testing, the ace68 was unable to advance into a demonstration neuron max due to the stretching on the distal shaft.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure in the internal carotid artery (ica) and m1 segment of the middle cerebral artery (mca) using a penumbra system ace 68 reperfusion catheter (ace68) and non-penumbra sheath.During the procedure, on the second pass under aspiration, the physician experienced slight resistance when pulling back the ace68.As the physician continued to pull back under aspiration, the ace68 stretched distally.The physician was able to remove the ace68.The procedure was completed using a new ace68 and the same sheath.There was no report of an adverse effect to the patient.
|
| |
|