| Model Number |
3183 |
| Medical Device Problem Code |
Low impedance (2285)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: date of event is unknown.Concomitant medical products and therapy dates: model: 3186, therapy date: unknown.
|
| |
|
Event or Problem Description
|
|
Device 2 of 2.Reference mfr.Report#: 1627487-2021-01309.It was reported the patient has been receiving ineffective therapy, overstimulation, and discomfort.An impedance check was taken and revealed low impedance on several contacts.Reprogramming was attempted but resolution was unsuccessful.In turn, the patient plans to undergo surgical intervention.
|
| |
|
Event or Problem Description
|
|
Device 2 of 2 reference mfr.Report#: (b)(4).Additional information obtained revealed surgical intervention took place on (b)(6) 2021.During the procedure the patient's leads were removed, cleaned, and reinserted into their ipg with impedances being within normal range.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
Device 2 of 2.Reference mfr.Report#: 1627487-2021-01309.
|
| |
|