• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OCTRODE LEAD KIT, 30CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OCTRODE LEAD KIT, 30CM LENGTH; SCS LEAD Back to Search Results
Model Number 3183
Medical Device Problem Code Low impedance (2285)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: date of event is unknown.Concomitant medical products and therapy dates: model: 3186, therapy date: unknown.
 
Event or Problem Description
Device 2 of 2.Reference mfr.Report#: 1627487-2021-01309.It was reported the patient has been receiving ineffective therapy, overstimulation, and discomfort.An impedance check was taken and revealed low impedance on several contacts.Reprogramming was attempted but resolution was unsuccessful.In turn, the patient plans to undergo surgical intervention.
 
Event or Problem Description
Device 2 of 2 reference mfr.Report#: (b)(4).Additional information obtained revealed surgical intervention took place on (b)(6) 2021.During the procedure the patient's leads were removed, cleaned, and reinserted into their ipg with impedances being within normal range.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Device 2 of 2.Reference mfr.Report#: 1627487-2021-01309.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 30CM LENGTH
Common Device Name
SCS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key11326107
Report Number1627487-2021-01310
Device Sequence Number728870
Product Code LGW
UDI-Device Identifier05414734401692
UDI-Public05414734401692
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2013
Device Model Number3183
Device Catalogue Number3183
Device Lot Number3339577
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/25/2021
Supplement Date Received by Manufacturer02/24/2021
03/17/2021
Initial Report FDA Received Date02/14/2021
Supplement Report FDA Received Date03/03/2021
03/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-