SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 7-8 9MM RT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
|
Back to Search Results |
|
| Model Number |
74035271 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, a jrny ii bcs art isrt trl sz 7-8 9mm rt was damaged.The device is broken and chipped, and also have some cracks.Incident was noticed during a tka while using instruments inside the patient.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A photo of the device is attached and confirms the failure mode.A crack is present on the device and pieces have broken off.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|