• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DRILL GUIDE HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DRILL GUIDE HANDLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674001
Device Problems Break (1069); Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference number: (b)(4).
 
Event Description
It was reported that, the bowed of a drill guide handle split and that pieces were falling/crumbling out.Incident occurred during a spd audit.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.Our investigation included a visual inspection of the device which confirms that the epoxy coating over the internal witness wire has degraded and has begun to come off.This failure mode has been previously identified.Since the manufacturing of the complaint device, design changes have been implemented to eliminate the occurrence of this failure.The device was manufactured in 1995.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, corrective and preventive action is indicated to mitigate future recurrence of similar events.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL GUIDE HANDLE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11326560
MDR Text Key231958855
Report Number1020279-2021-01305
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560469
UDI-Public03596010560469
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674001
Device Catalogue Number71674001
Device Lot Number489531 OR 5403321
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2021
Date Manufacturer Received01/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-