• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO EUPHORA RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO EUPHORA RX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number EUP2512X
Medical Device Problem Codes Burst Container or Vessel (1074); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Device Embedded In Tissue or Plaque (3165); Unspecified Tissue Injury (4559); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During a procedure an attempt was made to use a euphora rx balloon catheter to treat a severely calcified and severely tortuous lesion exhibiting 95% stenosis located in the mid left anterior descending artery (lad).The device was inspected with no issues noted.Negative prep was performed.The lesion was pre dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device.Excessive force was not used during delivery.The device was not kinked and restraightened during use.It was reported that the balloon ruptured occurred and upon inspection outside the patient body, it was noticed that portion of the balloon was detached.The physician believed that it was left in the target vessel.Patient was sent to cabg and had a heart open bypass.
 
Additional Manufacturer Narrative
Additional information: the device was being used to pre-dilate the lesion.The balloon ruptured on the third inflation.20 atm inflation pressure was applied prior to the balloon burst.The device was not moved or re-positioned while inflated.There was no resistance noted during withdrawal of the device.The detached portion of the device was not removed from the patient, no attempts were made to retrieve it.The open heart bypass was a direct result of this event.Patient was reported doing well and recovering after surgery.Correction: ime code medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Correction: b1, b2, h1, ime code imf code.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EUPHORA RX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
MDR Report Key11326998
Report Number9612164-2021-00635
Device Sequence Number19066071
Product Code LOX
UDI-Device Identifier00643169560017
UDI-Public00643169560017
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K143480
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/08/2022
Device Model NumberEUP2512X
Device Catalogue NumberEUP2512X
Device Lot Number220727967
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/11/2021
Supplement Date Received by Manufacturer02/17/2021
03/25/2021
Initial Report FDA Received Date02/15/2021
Supplement Report FDA Received Date03/12/2021
03/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient Weight95
-
-