• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PARTIAL TIBIAL CEMENTED SIZE C LEFT MEDIAL; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PARTIAL TIBIAL CEMENTED SIZE C LEFT MEDIAL; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994); Osteopenia/ Osteoporosis (2651)
Date of Event 02/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Foreign report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that patient underwent a procedure 21 days post implantation due to pain and a fractured tibia bone.The fracture was fixed with an implant plate.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This complaint was confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.X-rays were provided and reviewed by a health care professional.Review found periprosthetic fracture of the medial tibia with slight tibial implant subsidence.Bone quality appears osteopenic.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARTIAL TIBIAL CEMENTED SIZE C LEFT MEDIAL
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key11327762
Report Number3007963827-2021-00028
Device Sequence Number2660896
Product Code HSX
UDI-Device Identifier00880304812642
UDI-Public(01)00880304812642(17)300331(10)64468723
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42538000301
Device Lot Number64468723
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer03/25/2021
Initial Report FDA Received Date02/15/2021
Supplement Report FDA Received Date04/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
42558000201 64623123 42518200308 45029232
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
Patient Weight40
-
-