| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's system pcb assembly to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Physio-control further evaluated the removed system pcb assembly and determined that the cause of the reported issue was due to a malfunctioning crystal, designator y1, on the single board computer.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device is constantly rebooting itself.There was no report of patient use associated with the reported event.
|
| |
|