| Model Number |
1010-11-120 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Fall (1848); Bone Fracture(s) (1870); Inadequate Osseointegration (2646)
|
| Date of Event |
01/28/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient fell and developed a femoral fracture distal to the implanted actis size 12 stem.Surgeon removed actis stem and head.Pinnacle cup and liner remained implanted.It was also indicated that there was loosening of the stem at the bone to implant interface.Doi: (b)(6) 2018; dor: (b)(6) 2021; affected side: left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|