| Model Number |
LTV 1200 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The equipment was received in vyaire facility for evaluation.Service technician was not able to reproduce the reported problem "set bpm auto cycles".The ventilator passed 96 hours extended tests at customer settings without any unusual alarms or conditions.However, the ventilator failed troubleshooting final test at pressure and oxygen blending performance.Bench testing revealed o2 blender is creating the non-conformance.Service technician replaced o2 blender with a good known test o2 blender and ventilator passed pressure and oxygen blending performance.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported to vyaire medical that the ltv (lap top ventilator) 1200 shut off while on a patient and experienced ventilator inoperable.The customer confirmed that there was no patient harm associated with the reported event.
|
| |
|