Model Number LTV 1200
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The equipment was received in vyaire facility for evaluation.Service technician was not able to reproduce the reported problem "set bpm auto cycles".The ventilator passed 96 hours extended tests at customer settings without any unusual alarms or conditions.However, the ventilator failed troubleshooting final test at pressure and oxygen blending performance.Bench testing revealed o2 blender is creating the non-conformance.Service technician replaced o2 blender with a good known test o2 blender and ventilator passed pressure and oxygen blending performance.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
The customer reported to vyaire medical that the ltv (lap top ventilator) 1200 shut off while on a patient and experienced ventilator inoperable.The customer confirmed that there was no patient harm associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LTV 1200 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
MDR Report Key11328836
Report Number2021710-2021-13378
Device Sequence Number10996656
Product Code CBK
UDI-Device Identifier00845873002696
UDI-Public(01)00845873002696(11)20150506
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K060647
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLTV 1200
Device Catalogue Number18888-201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2021
Initial Report FDA Received Date02/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-