Model Number BELLAVISTA 1000E |
Device Problem
Smoking (1585)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/18/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.As a resolution, main electronics, power supply and power board will be replaced.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
It is reported to vyaire medical that the bellavista 1000 experienced a burning/smoke arising from the device that caused damaged to the components.At this time, there is no information regarding patient involvement associated with the reported event.
|
|
Manufacturer Narrative
|
Result of investigation: vyaire medical was able to verify the customer's reported issue.Bench testing revealed no 24v output of the ac/dc power supply.Damaged components visible on the board.The reported issue is most likely attributed to fluctuations within the power supply.The defective component was replaced and the issue was resolved.
|
|