• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number BELLAVISTA 1000E
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2021
Event Type  malfunction  
Manufacturer Narrative
Vyaire file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.As a resolution, main electronics, power supply and power board will be replaced.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
It is reported to vyaire medical that the bellavista 1000 experienced a burning/smoke arising from the device that caused damaged to the components.At this time, there is no information regarding patient involvement associated with the reported event.
 
Manufacturer Narrative
Result of investigation: vyaire medical was able to verify the customer's reported issue.Bench testing revealed no 24v output of the ac/dc power supply.Damaged components visible on the board.The reported issue is most likely attributed to fluctuations within the power supply.The defective component was replaced and the issue was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BELLAVISTA
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
IMTMEDICAL AG
gewerbestrasse 8
buchs, 9470
SZ  9470
MDR Report Key11329766
MDR Text Key231929136
Report Number3004553423-2021-00850
Device Sequence Number1
Product Code CBK
UDI-Device Identifier07640149380019
UDI-Public(01)07640149380019
Combination Product (y/n)N
PMA/PMN Number
K163127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBELLAVISTA 1000E
Device Catalogue Number301.100.000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2021
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/15/2021
Supplement Dates Manufacturer Received08/26/2021
Supplement Dates FDA Received09/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-