• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7133
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The packaging, shaft, hypotube, tip and balloon were microscopically and visually examined.The pouch was opened along the vendor seal.There were numerous kinks.There was a complete separation at 20.3cm distal of the strain relief.There was contrast and blood in the inflation lumen, and blood in the guidewire lumen.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
Reportable based on device analysis completed on 21-jan-2021 it was reported that the packaging was damaged.A percutaneous coronary intervention was being performed.A 1.50mm x 15mm emerge balloon catheter was selected for use.However, the physician found that the package was damaged.There were no patient complications reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11329800
Report Number2134265-2021-00927
Device Sequence Number1045946
Product Code LOX
UDI-Device Identifier08714729806189
UDI-Public08714729806189
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K121196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/19/2022
Device Model Number7133
Device Catalogue Number7133
Device Lot Number0025464530
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/21/2021
Initial Report FDA Received Date02/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient Weight66
-
-