| Model Number |
VENUL14140 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.However, images were provided for review.The investigation of the reported event is currently underway.Expiry date: 10/2022.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement through right popliteal vein, the device allegedly failed to expand.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement through right popliteal vein, the device allegedly failed to expand.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: a voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.H10: manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the physical sample was not available.Images provided demonstrate that the stent is visibly constricted in the area of the silicone stent brake.Distal and proximal of that section the stent is regularly expanded so that it appears like an hour-glass which indicates that the deployment sheath had been retracted and that the stent adheres to the silicone brake, which leads to confirmed result for expansion issue.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 10/2022), g3.H11: h6 (method, result).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|