MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION POLYFLUX 170H; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION POLYFLUX 170H; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 103579
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that upon starting treatment with a polyflux 170h, an external dialysate leak was observed at the header.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the device was received for evaluation.The visual inspection by naked eye of the product showed the product wet.A leak test of the dialysate side was performed and a leak was observed.The cause was a crack in the welding zone.The reported condition was verified.The cause of the crack could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 170H
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11331340
Report Number9611369-2021-00022
Device Sequence Number883299
Product Code KDI
UDI-Device Identifier07332414037932
UDI-Public(01)07332414037932
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
K043342
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Catalogue Number103579
Device Lot Number0-4614-H-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2021
Initial Date Received by Manufacturer 01/25/2021
Supplement Date Received by Manufacturer02/22/2021
Initial Report FDA Received Date02/16/2021
Supplement Report FDA Received Date03/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-