• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX ST100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX ST100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115959
Device Problems Break (1069); Fluid/Blood Leak (1250); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that approximately fifty hours into patient treatment with a prismaflex st100 set, the "border part" of the drain bag was observed to be damaged and an external fluid leak occurred.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information added to h6 and h10 h10: the actual device was not available for evaluation.However, based on the information provided, the same bag was used for 50 hours.This is a misuse of the effluent bag.During the investigation of similar complaints it has been noticed that the leak on the drain bags occurred mainly after several hours of treatment, after being filled and emptied several times.The filling/emptying cycles generates physical constraints on the bags, near the welding of the exhausting tube, eventually causing pinholes to form and leakage to happen.However, the prismaflex control unit operator's manual indicates to ¿change fluid bags when the appropriate caution alarm occurs (pbp bag empty, replacement bag empty, dialysate bag empty or effluent bag full)¿.The prismaflex st100 set and all the accessories provided within the set, including the drain bag, is a single use product.This means that once used the product must not be reused (i.E.Emptied and reused during the same therapy) and should be discarded.The prismaflex control unit warns the operator when the effluent bag is full and it is specified in the on-screen instruction delivered by the prismaflex control unit that the drain bag should be disconnected and changed for a new bag.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX ST100
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11331381
MDR Text Key242767651
Report Number8010182-2021-00053
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number115959
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received03/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PRISMAFLEX MACHINE
-
-