According to the customer report, presence of water/moisture spill was confirmed in the ventilator's gas module.The ventilator could be used again after drying the gas module.No parts were replaced and no ventilator logs were provided.Water/moisture in supply gases into a gas modules can cause failure which might lead to a stop of ventilation or high pressures.Alarms will be generated.According to the user's manual, maximum levels of water, (h2o < 7 g/m3) in the supplied gases to the ventilator must not be exceeded.The most probable source of moisture/water into a gas module is moisture from the hospital's gas supply system and/or external compressor.The root cause of the reported event has not been determined.Based on the above information this complaint will be closed.
|