| Model Number |
IPN000325 |
| Medical Device Problem Code |
Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that when the user attempted to start the intra-aortic balloon pump (iabp) with using the intra-aortic balloon (iab), a purge alarm occurred and the iabp didn't start working.As a result, a back-up iabp was used without any issues.There was no report of patient complications, serious injury or death.
|
| |
|
Event or Problem Description
|
|
It was reported that when the user attempted to start the intra-aortic balloon pump (iabp) with using the intra-aortic balloon (iab), a purge alarm occurred and the iabp didn't start working.As a result, a back-up iabp was used without any issues.There was no report of patient complications, serious injury or death.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).No iabp part or recorder strip was returned to teleflex chelmsford for investigation.The reported complaint of purge failure alarm is not able to be confirmed.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
|
| |
|