MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT EZ; BLOOD GLUCOSE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT EZ; BLOOD GLUCOSE METER Back to Search Results
Model Number 7252
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
 
Event or Problem Description
The customer reported that one of his blood glucose readings was not stored in the memory of the contour next ez meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
The customer retuned the suspected contour next ez meter and contour next test strips for evaluation.There were only four strips in the returned vial of test strips.Therefore, the returned meter was tested with the returned test strips and in-house contour next test strips using a blood sample, which gave a satisfactory performance.The blood glucose readings obtained during in-house testing were properly stored in the meter's memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT EZ
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
MDR Report Key11333224
Report Number1810909-2021-00089
Device Sequence Number1034863
Product Code NBW
UDI-Device Identifier00301937252013
UDI-Public00301937252013
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111268
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number7252
Device Catalogue Number7252
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2021
Initial Date Received by Manufacturer 01/19/2021
Supplement Date Received by Manufacturer04/08/2021
Initial Report FDA Received Date02/16/2021
Supplement Report FDA Received Date04/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-