• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT CREA CARTRIDGE; CREA+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT CREA CARTRIDGE; CREA+ CARTRIDGE Back to Search Results
Catalog Number 03P84-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Pain (1994)
Date of Event 01/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident# (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event or Problem Description
On 01-feb-2021, abbott point of care was contacted by a customer regarding i-stat crea cartridges that yielded a suspected discrepant creatinine result of 2.8 mg/dl on a (b)(6) year old female patient presented with pain, post-op hysterectomy.There was no patient information available at the time of this report.Return product is not available for investigation.Method:i-stat.Date: (b)(6) 2021.Collected: 1530.Tested: 153.Result: 2.8 mg/dl.Sample:venous.Method : (b)(6).Date : (b)(6) 2021.Collected : ni.Tested :ni.Result : 1.2 mg/dl.Sample: ni.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Event or Problem Description
Na.
 
Additional Manufacturer Narrative
Apoc incident: (b)(4).The investigation was completed on (b)(6) 2021.A review of the device history record confirmed the lot passed finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Af, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency has been identified for crea cartridge lot a20278.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CREA CARTRIDGE
Common Device Name
CREA+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key11334305
Report Number2245578-2021-00009
Device Sequence Number2671829
Product Code CGL
UDI-Device Identifier10054749000125
UDI-Public10054749000125
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K973292
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/02/2021
Device Catalogue Number03P84-25
Device Lot NumberA20278
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/01/2021
Supplement Date Received by Manufacturer03/01/2021
Initial Report FDA Received Date02/16/2021
Supplement Report FDA Received Date03/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-