• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIG RP AOX CRV INS SZ3 10; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIG RP AOX CRV INS SZ3 10; SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT Back to Search Results
Model Number 1961-92-031
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/21/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical adverse event received for anterior scar development, patellar grinding, pain.Event is serious and is considered moderate.Event is possibly related to device and is probably related to procedure.Date of implantation: (b)(6) 2019, date of event (onset): (b)(6) 2021, date of intervention: (b)(6) 2021, (left knee).Treatment: arthroscopic debridement of adhesions.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG RP AOX CRV INS SZ3 10
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11334533
Report Number1818910-2021-03111
Device Sequence Number16809473
Product Code NJL
UDI-Device Identifier10603295076162
UDI-Public10603295076162
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1961-92-031
Device Catalogue Number196192031
Device Lot Number9113991
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/05/2021
Supplement Date Received by Manufacturer03/25/2021
Initial Report FDA Received Date02/16/2021
Supplement Report FDA Received Date03/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPUY CMW 2 20G; DEPUY CMW 2 20G; MBT CEM KEEL TIB TRAY SZ3; PFC*SIGMA/OV/DOME PAT 3PEG,35; SIG C/R POROCOAT FEM LT SZ 3; SIG RP AOX CRV INS SZ3 10; DEPUY CMW 2 20G; DEPUY CMW 2 20G; MBT CEM KEEL TIB TRAY SZ3; PFC*SIGMA/OV/DOME PAT 3PEG,35; SIG C/R POROCOAT FEM LT SZ 3; SIG RP AOX CRV INS SZ3 10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient Weight79
-
-